Open original FDA record ↗Recall record
- FDA identifier
- Z-1835-2020
- Company in FDA record
- ICU Medical, Inc.
- FDA record creation date
- May 1, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm identified several software issues through internal testing of its Cogent Hemodynamic Monitoring System. These issues may result in inaccurate monitoring data, unit shutdown, and/or loss of display of a secondary parameter.
Source facts
- Action
- Customers were notified of the recall via UPS letter on 7/12/2019. The firm is voluntarily notifying customers and will contact customers to schedule a software update once the updated software is available. The firm has instructed customers to use the mitigation instructions detailed in the customer letter to reduce the likelihood of the software conditions occurring. The firm also requested customers complete a response form and return it to the firm. For additional information or technical assistance contact ICU Medical at 1-866-829-9025, option 4.
- Initiated
- 2019-07-12
- Root cause
- Under Investigation by firm
- Terminated
- 2023-05-10
- Distribution
- US Nationwide Distribution in the states of FL, WI, NY, NC, VT, IL, MD, CA, NE, WA OUS: None
- Recall number
- Z-1835-2020
- Recall status
- Terminated
- CDRH record ID
- 180463
- FDA record created
- 2020-05-01
- Recall event number
- 85300