Skip to content

Public FDA record

Z-1836-2013: Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable

Recall record Z-1836-2013 for St. Jude Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1836-2013
Company in FDA record
St. Jude Medical
FDA classification date
Jul 31, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne

Source facts

Action
St. Jude Medical sent a Important Medical Device Recall Advisory dated July 26, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. To further assist in your patient care we are providing you with a list of all serial numbers we show have been distributed to you. ( See Attachment A ). Further questions please call (972) 309-2154.
Initiated
2012-07-26
Root cause
Device Design
Terminated
2015-08-21
Distribution
Worldwide Distribution--USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Belgium, Brazil, canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Japan, Mexico, Netherlands, Saudi, South Africa, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Recall number
Z-1836-2013
Recall status
Terminated
CDRH record ID
112563
FDA classified
2013-07-31
Recall event number
63108

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗