Skip to content

Public FDA record

Z-1836-2020: ABL800 FLEX model number 393-800 and 393-801, UDI:(01)05700693938004

Recall record Z-1836-2020 for Radiometer Medical ApS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1836-2020
Company in FDA record
Radiometer Medical ApS
FDA record creation date
May 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from cyanosis and low blood pressure to cardiac arrhythmias and severe impairment of cardiac function.

Source facts

Action
On 3/5/2020 a "Urgent: Medical device Recall" letter was hand delivered to effective consignees. In addition to providing information on the recall products, the customer notification as consignees to take the following actions: Solution provided by Radiometer: 1. Your Radiometer representative will exchange the power supply of your ABL800 analyzer at the earliest convenience. Your Actions: 1. Cease using the affected ABL800 analyzer and put it on quarantine. 2. Shut down the affected ABL800 analyzer and disconnect the power cord. 3. Complete the Recall Response Form (last page of this letter) and submit to your Radiometer representative 4. If you have any questions, please contact us at 1-800-736-0600 option 1, Monday Friday 9am EST to 7pm EST.
Initiated
2020-03-06
Root cause
Component design/selection
Distribution
US: in the state of California
Recall number
Z-1836-2020
Recall status
Open, Classified
CDRH record ID
180206
FDA record created
2020-05-01
Recall event number
85166

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗