Z-1837-2014: G7 Acetabular System Positioning Guide Rod, orthopedic surgical instrument for hip prosthesis PN 11000345

Recall record Z-1837-2014 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1837-2014
Company in FDA record
Biomet, Inc.
FDA classification date
Jun 19, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Guide rod may fracture during impaction of the acetabular shell and be retained in the inserter or the patient and could result in a delay in surgery.

Source facts

Action
The firm issued the following notice on May 7, 2014: URGENT MEDICAL DEVICE RECALL NOTICE: This notification is to inform you of an Urgent Medical Device Recall initiated by Biomet Orthopedics ("Biomet") which involves Part Numbers: 110003458 G7 Positioning Guide Rod and 110003500 G7 Positioning Guide Post. These instruments have been consigned and/or invoiced to your account. Biomet has initiated this action following an investigation which identified that positioning guide rod may fracture during impaction of the acetabula shell. If the positioning guide rod fractures, then the positioning guide post may become stuck to the insertion handle. There is no adverse health outcome expected for the patient. This action requires the immediate location and discontinued use of the product and its return to Biomet. The following actions are REQUIRED: Immediately locate and remove the identified device(s) listed below from circulation. Carefully follow the instructions on the enclosed "FAX Back Response Form. Fax a copy of the Response Form to 574-372-1683 prior to return of product. Use priority carrier for your shipment. If you have further distributed this product, you MUST notify hospital personnel of this action via the enclosed "Dear Risk/Recall Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices. However, you are charged with the location and return of these products.
Initiated
2014-05-07
Root cause
Use error
Terminated
2015-10-30
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Netherlands, New Zealand, and Japan.
Recall number
Z-1837-2014
Recall status
Terminated
CDRH record ID
127364
FDA classified
2014-06-19
Recall event number
68310

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗