At a PROBEAT III site installed in Japan, the operator of the system found that the radiation was delivered inconsistent with the treatment plan and stopped the irradiation. The irradiation control system may be potentially affected when rebooting a part of the control system.
Source facts
Action
The recalling firm notified the one consignee about the incident that occurred in Japan on 4/7/15, then gave them an update on 4/13/15 and 4/21/15. The written notification had details for the consignee on how to prevent the device from malfunctioning. The recalling firm also had held conversations with the consignee to discuss the issue and revised work instructions via telephone conference and on site Hitachi personnel.
Initiated
2015-04-21
Root cause
Device Design
Terminated
2016-03-24
Distribution
Distribution in US (Texas) and Japan
Recall number
Z-1838-2015
Recall status
Terminated
CDRH record ID
135952
FDA classified
2015-07-18
Recall event number
71056