Skip to content

Public FDA record

Z-1839-2019: LVivo EF app on Vscan Extend. Sold under the following product names: a. Vscan Extend Dual DICOM, Mod

Recall record Z-1839-2019 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1839-2019
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Jun 18, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Overestimation bias in automatically calculated ejection fraction (EF) values while using LVivo EF app on the Vscan Extend product.

Source facts

Action
GE Healthcare notified customers on about 05/21/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were instructed to discontinue use of the LVivo EF app until a correction is available and installed on the Vscan Extend. However Vscan Extend can continue to be used for all other existing cleared indications including all other Vscan Extend apps and measurements. GE Healthcare will correct all affected products at no cost. When the LVivo EF app update is available through GE Marketplace, customers will receive an email notification sent to the email address they set up during the Vscan Extend device registration process. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2019-05-21
Root cause
Software design
Distribution
Worldwide distribution - US nationwide distributed in the states of CA, FL, IL, MA , MD, MI, MN, NY, SC, TX, UT. Countries of Australia, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Korea (Republic Of), Netherlands, Romania, Sweden, Switzerland, Thailand, United Kingdom.
Recall number
Z-1839-2019
Recall status
Open, Classified
CDRH record ID
173162
FDA record created
2019-06-18
Recall event number
82938

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗