Skip to content

Public FDA record

Z-1840-2019: LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYST

Recall record Z-1840-2019 for CooperSurgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1840-2019
Company in FDA record
CooperSurgical, Inc.
FDA record creation date
Jun 19, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.

Source facts

Action
On April 23, 2019, the firm distributed Urgent Medical Device Safety Alert letters to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were asked to immediately discontinue use of the discrepant product, quarantine the product, and replace the 2 main fuses with the 2 fuses provided with the firm's mailing. Customers were provided with instructions to replace the fuses. Once customers have successfully completed the fuse replacement, they are asked to complete the Acknowledgement and Receipt form and return it to CooperSurgical. If you have any questions, please contact the firm at 203-601-5200 ext. 3300.
Initiated
2019-04-23
Root cause
Process control
Terminated
2020-04-24
Distribution
US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI.
Recall number
Z-1840-2019
Recall status
Terminated
CDRH record ID
172775
FDA record created
2019-06-19
Recall event number
82821

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.