Open original FDA record ↗Recall record
- FDA identifier
- Z-1840-2019
- Company in FDA record
- CooperSurgical, Inc.
- FDA record creation date
- Jun 19, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.
Source facts
- Action
- On April 23, 2019, the firm distributed Urgent Medical Device Safety Alert letters to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were asked to immediately discontinue use of the discrepant product, quarantine the product, and replace the 2 main fuses with the 2 fuses provided with the firm's mailing. Customers were provided with instructions to replace the fuses. Once customers have successfully completed the fuse replacement, they are asked to complete the Acknowledgement and Receipt form and return it to CooperSurgical. If you have any questions, please contact the firm at 203-601-5200 ext. 3300.
- Initiated
- 2019-04-23
- Root cause
- Process control
- Terminated
- 2020-04-24
- Distribution
- US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI.
- Recall number
- Z-1840-2019
- Recall status
- Terminated
- CDRH record ID
- 172775
- FDA record created
- 2019-06-19
- Recall event number
- 82821