Z-1840-2021: Myriad-LX Light Source-for delivery of light to the surgical field to enhance visualization of tissue dur

Recall record Z-1840-2021 for Nico Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1840-2021
Company in FDA record
Nico Corp.
FDA record creation date
Jun 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient or end user.

Source facts

Action
The Firm, Nico Corporation, issued "URGENT: MEDICAL DEVICE RECALL" Letter dated 4/30/21 via email on 4/30/21. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY YOU Illumination Pack Field Correction 1. Determine the current number of LX Illumination Packs at your facility with the lot numbers referenced above. 2. Locate the electronic version of the updated Instructions-for-Use for LX Illumination Packs that was provided with this letter via email (document number LL-5697 03). 3. Print the updated Instructions-for-Use in a quantity sufficient to cover all LX Illumination Packs at your facility with the lot numbers referenced above. 4. Replace the existing Instructions-for-Use (LL-5697_01) with the updated Instructions for-Use (LL-5697 03). 5. Discard the existing Instructions-for-Use (LL-5697_01). LX Light Source Removal 1. Your local NICO representative will notify when he/she will be on site to deliver the return _kit for retrieval of your LX Light Source. 2. Upon receipt of the return kit, remove your LX Light Source from use and package it using the materials provided within the kit. 3. Tape the box closed, affix the prepaid shipping label and deliver it to your shipping department. 4. Upon receipt of your LX Light Source, NICO will modify it and will then return it to your local NICO representative. He/she will then coordinate with you to return it to your facility. 5. Upon receipt of your modified LX Light Source, remove it from its packaging and begin using it again. 6. Locate and print the Response Form that was provided with this letter via email. 7. Complete the Response Form (including wet signature) and email it to jay.dittman@niconeuro.com. A scanned version or a photo of the Response Form is acceptable. NICO has opened a corrective action to address this issue. If you have any questions or need additional information, please contact me jay.dittman@niconeuro.com or 317-660-7118 ext. 204
Initiated
2021-04-30
Root cause
Device Design
Terminated
2024-02-05
Distribution
US Nationwide Distribution
Recall number
Z-1840-2021
Recall status
Terminated
CDRH record ID
187291
FDA record created
2021-06-09
Recall event number
87882

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗