Z-1841-2014: Integra UCR 3mm Hex, Large Axial Driver. The UCR Hex Axial Driver is a surgical instrument used to imp

Recall record Z-1841-2014 for SeaSpine Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1841-2014
Company in FDA record
SeaSpine Inc
FDA classification date
Jun 20, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Integra has identified through an investigation of complaints that there may be the potential for the tip of a single lot number of the UCR 3mm Hex, Large Axial Driver, 91-1123 to twist or become rounded during use.

Source facts

Action
Integra sent an Urgent Voluntary Medical Device Recall letter dated May 21, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: 1. If you have distributed the affected product, identify those customers and forward the attached Integra Medical Device Recall Notice to them. 2. Inspect your inventory of UCR 3mm Hex, Large Axial Drivers (Catalogue Number: 91- 1123). If you have product with lot number MO16090B, STOP USING IMMEDIATELY. If you have product of any other lot number, you can continue to use it. The event has been isolated to this single lot. 3. We also request that regardless if you have a UCR Driver with the affected lot, you complete the attached form and return it to us as soon as possible as indicated. We recommend you keep a copy of the form for your records. 4. When your form is received, Customer Service will contact you and provide an RMA number, directions to return the product, and input an order to replace the quantity you indicate on the form. Should you have any questions regarding these instructions, please contact Customer Service at 1-866- 942-8698. We apologize for any inconvenience this may cause and thank you for your cooperation in this effort. : .
Initiated
2014-05-21
Root cause
Under Investigation by firm
Terminated
2014-12-15
Distribution
US Distribution including the states of : AL, CA, OH, NM and TN.
Recall number
Z-1841-2014
Recall status
Terminated
CDRH record ID
127469
FDA classified
2014-06-20
Recall event number
68362

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗