Open original FDA record ↗Recall record
- FDA identifier
- Z-1841-2016
- Company in FDA record
- Integra York PA, Inc
- FDA classification date
- May 31, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One lot of self-lock chisel blade handles have a wider slot that holds the chisel blade that may result in the inability to firmly secure the blade in the handle.
Source facts
- Action
- Integra mailed out an Urgent: Voluntary Medical Device Recall letter to customers informing them to evaluate their inventory for the affected product. If product was shipped to customers, they were asked to identify their customers and to stop shipping immediately. Integra requested that customers complete the attached form that was sent with the notification indicating whether or not they had the affected product on hand and customer service will contact them to provide a Return Authorization number for the product's return. Customers could reach Customer Service at 866-854-8300.
- Initiated
- 2016-03-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-04-17
- Distribution
- Worldwide distribution. US nationwide, Mexico, and Korea.
- Recall number
- Z-1841-2016
- Recall status
- Terminated
- CDRH record ID
- 146227
- FDA classified
- 2016-05-31
- Recall event number
- 74079