Open original FDA record ↗Recall record
- FDA identifier
- Z-1842-2016
- Company in FDA record
- ConMed Corporation
- FDA classification date
- May 31, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Metal shavings released from burs during use are due to contact between the bur and the hood of the PreBent Bur assembly. During the manufacturing process an assembly error occurred which caused inadvertent mixing of two sizes of bearings.
Source facts
- Action
- ConMed sent an Urgent Medical Device Recall Notification letter dated December 17, 2015, to all affected customers. Customers are urged to stop use and return these devices immediately. Customers who have further distributed these items are urged to contact all of those organizations immediately and provide a copy of the recall notification letter. It is imperative that all end users of these devices receive the recall notice and respond immediately. Customers with questions can call 315-624-3237 or email largohpshb@conmed.com. The devices should be returned to ConMed Corporation NCR-47057 11155 Concept Blvd. Largo, FL 33773. For questions regarding this recall call 315- 624-3533.
- Initiated
- 2015-12-21
- Root cause
- Mixed-up of materials/components
- Terminated
- 2018-05-24
- Distribution
- Worldwide Distribution - Nationwide to CA, CO, FL, IN, MD, MN, NC, NY, OH, TX, UT & VA; International: Canada, Wales
- Recall number
- Z-1842-2016
- Recall status
- Terminated
- CDRH record ID
- 143103
- FDA classified
- 2016-05-31
- Recall event number
- 73123