Open original FDA record ↗Recall record
- FDA identifier
- Z-1844-2019
- Company in FDA record
- Cook Inc.
- FDA classification date
- Jun 19, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
may contain a damaged gray safety lock knob, which could potentially result in difficulty or an inability to fully deploy the graft via the standard or troubleshooting method provided in the Instructions for Use (IFU). Potential adverse events that may occur if an affected product is used include a prolonged procedure and open surgical intervention.
Source facts
- Action
- Urgent Medical Device Recall notification letters dated 4/26/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. For questions or concerns, contact Cook Medical Customer Relations at 800-457-4500 or 812-339-2235.
- Initiated
- 2019-04-26
- Root cause
- Under Investigation by firm
- Terminated
- 2019-11-08
- Distribution
- No US distribution. Product distributed to the following foreign countries: Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile Columbia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
- Recall number
- Z-1844-2019
- Recall status
- Terminated
- CDRH record ID
- 173095
- FDA classified
- 2019-06-19
- Recall event number
- 82904