Skip to content

Public FDA record

Z-1844-2019: Zenith Alpha Abdominal Endovascular Graft, Reference Part Numbers: ZIMB-22-108, ZIMB-22-118, ZIMB-22-128,

Recall record Z-1844-2019 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1844-2019
Company in FDA record
Cook Inc.
FDA classification date
Jun 19, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

may contain a damaged gray safety lock knob, which could potentially result in difficulty or an inability to fully deploy the graft via the standard or troubleshooting method provided in the Instructions for Use (IFU). Potential adverse events that may occur if an affected product is used include a prolonged procedure and open surgical intervention.

Source facts

Action
Urgent Medical Device Recall notification letters dated 4/26/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. For questions or concerns, contact Cook Medical Customer Relations at 800-457-4500 or 812-339-2235.
Initiated
2019-04-26
Root cause
Under Investigation by firm
Terminated
2019-11-08
Distribution
No US distribution. Product distributed to the following foreign countries: Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile Columbia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Recall number
Z-1844-2019
Recall status
Terminated
CDRH record ID
173095
FDA classified
2019-06-19
Recall event number
82904

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗