Open original FDA record ↗Recall record
- FDA identifier
- Z-1845-2011
- Company in FDA record
- BD Biosciences, Systems & Reagents
- FDA classification date
- Mar 30, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Results and gates could be mismatched if you utilize the undo function to undo a gate that was deleted.
Source facts
- Action
- Important Product Information letters, dated September 2007, were sent the week of 9/14/2001 via Federal Express to all affected customers. The letters identified the affected product along with the reason for recall. The letter states that "as long as customers do not use the Undo function to undo a gate deletion, the software can be used with full functionality." The letter also states that BD will change the issue in the next software version. If customers believe that they may have analyzed sample files by utilizing the Undo function, then it is recommended that the sample files be re-analyzed. If customers need more information they should contact the Customer Support Center at 877-232-8995.
- Initiated
- 2007-09-07
- Root cause
- Software design
- Terminated
- 2011-03-31
- Distribution
- Worldwide Distribution
- Recall number
- Z-1845-2011
- Recall status
- Terminated
- CDRH record ID
- 96797
- FDA classified
- 2011-03-30
- Recall event number
- 57613