Open original FDA record ↗Recall record
- FDA identifier
- Z-1845-2016
- Company in FDA record
- Olympus Corporation of the Americas
- FDA record creation date
- May 31, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
New reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they plan to implement as well as a new cleaning brush (MAJ-1534), which is enclosed with the notification.
Source facts
- Action
- Olympus notified customers on 03/14/2016 via an Urgent Safety Notification letter that referenced discontinuation of previously distributed copies of the TJF-160F/VF Reprocessing Manual. Customers were asked to inspect their inventory and discard any existing inventory of TJF-160F/VF Reprocessing Manuals. They were asked to update their information with the enclosed TJF-160F/VF Reprocessing Manual, which contains new brushing and flushing steps for the TJF-160F/VF elevator mechanisms and the forceps elevator recesses. Instuctions for cleaning and a new cleaning brush (used for cleaning the device) was included with the notification. Customers were asked to ensure all reprocessing personnel are completely knowledgeable and throughly training on the new reprocessing instructions in the new Reprocessing Manual.
- Initiated
- 2016-03-15
- Root cause
- Under Investigation by firm
- Terminated
- 2018-02-06
- Distribution
- Nationwide:
- Recall number
- Z-1845-2016
- Recall status
- Terminated
- CDRH record ID
- 144647
- FDA record created
- 2016-05-31
- Recall event number
- 73640