There exists a possible position sensor fault in the swivel base axis not being detected by the system software. When moving, the system could potentially exceed its usual speed, resulting in a collision. No injuries reported.
Source facts
Action
Siemens sent an Important Customer Safety Notice letter dated May 22, 2015, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter also informed customers that Siemens would contact them to schedule the installation of a software update to resolve the problem.
Initiated
2015-05-22
Root cause
Software design
Terminated
2016-12-13
Distribution
US Nationwide Distribution
Recall number
Z-1846-2015
Recall status
Terminated
CDRH record ID
137435
FDA classified
2015-06-25
Recall event number
71359