Z-1846-2019: Giraffe Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neo

Recall record Z-1846-2019 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1846-2019
Company in FDA record
GE Healthcare, LLC
FDA classification date
Jul 11, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.

Source facts

Action
GE Healthcare notified customers on about 03/15/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were instructed to 1) Inspect the warmer bedside panels latch areas and the areas connecting the panel to the bed for any cracks or damages per the addendum included in the mailing. If any portion of the bedside panels or latches are cracked or broken, stop use of the warmer. Complete and return the attached response form in via FAX TO: +1-410-630-5579, or scanning or taking a photo of the completed form e-mailing to: MIC.Recall@ge.com indicating what components require replacement. If the Warmer bedside panels or latches have no cracks or damage, continue to Safety Instruction 2. 2) If the Warmer bedside panels or latches have no cracks or damage, you may continue to use your warmer. Apply Safety labels included in the mailing to the warmer by following the below instructions: a) Clean area with a cleaning solution recommended in the Operation and Maintenance Manual. b)Wait for the area to dry. c)Peel off label 1 and apply it to the front (south) bedside panel at the location shown in the illustration. d)Peel off label 2 and apply it to the left (west) bedside panel at the location shown in the illustration. e)Peel off label 3 and apply it to the right (east) bedside panel at the location shown in the illustration. 3) Review the attached Addendum and place it with your Warmer Operation and Maintenance Manual. Customers were instructed that GE Healthcare will replace all cracked or damaged bedside panels/latches at not cost. Complete and return the attached "Customer Response" form via e-ail to MIC.Recall@ge.com and a GE Representative contact you to arrange for a replacement. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative. On about 06/14/2019, GE Healthcare issued a press release and sent customers an "AMENDED - URGENT MEDICAL DEVICE RECALL" letter.
Initiated
2019-03-15
Root cause
Device Design
Terminated
2024-05-21
Distribution
Worldwide Distribution: US (Nationwide distribution) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Guam, Northern Mariana Islands; and to countries of: Albania, Algeria, Argentina, Aruba, Australia, , Austria, Azerbaijan, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Bolivia, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guam, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Korea (Republic of), Kuwait, Lebanon, Libya, Lithuania, Macau, Macedonia, Malaysia, Maldives, Mauritius, Mexico, Moldova, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, STATE OF, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Recall number
Z-1846-2019
Recall status
Terminated
CDRH record ID
171685
FDA classified
2019-07-11
Recall event number
82485

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗