Recall record Z-1846-2020 for Applied Medical Resources Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1846-2020
Company in FDA record
Applied Medical Resources Corp
FDA classification date
May 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.

Source facts

Action
On November 08, 2019, a "Urgent: Medical Device Recall" letter was sent via UPS to all affected customers. In addition to providing information on the recalled products, the letter asked customers to do the following: 1. Check your inventory for the recalled product. 2. Complete the Recall Notification Confirmation Form to acknowledge the recall and indicate if your facility is returning or has already used product from the lot listed above. 3. If no product is being returned, please indicate on the Recall Notification Confirmation Form. 4. If you are a distributor, please notify any facilities to which you distributed the affected product. Please also complete Page 4 of the Recall Notification Confirmation Form. 5. Return the Recall Notification Confirmation Form to Applied Medical by email to recall60810330@appliedmedical.com or faxing it to (949) 713-8908. 6. Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical (Product Return Instructions are on Page 5). 7. For product return questions, please contact Jaclene Rios-Simpson, Senior Manager, Customer Relations at (949) 713-8688 or jrios@appliedmedical.com. 8. For regulatory questions, please contact me, Lauren Contursi, at (949) 713-8767 or lcontursi@appliedmedical.com.
Initiated
2019-10-24
Root cause
Component change control
Terminated
2023-06-01
Distribution
Worldwide distribution - United States: Nationwide OUS: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Poland, Saudi Arabia, South Africa, Spain, Taiwan, United Kingdom
Recall number
Z-1846-2020
Recall status
Terminated
CDRH record ID
179984
FDA classified
2020-05-07
Recall event number
85057

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗