Z-1848-2011: HEMOCUE AB GLUCOSE 201 DM ANALYZER, in vitro diagnostic device for quantitative determination of blood

Recall record Z-1848-2011 for Hemo Cue, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1848-2011
Company in FDA record
Hemo Cue, Inc.
FDA classification date
Mar 30, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Incorrect configurations settings in software. As a result, certain error codes do not trigger appropriately. These error codes are: E01, E02 and E05 which are all related to the optronic performance of the analyzer. Possible incorrect results.

Source facts

Action
Hemocue AB sent Customer Notification "Notice of Medical Device Correction on HemoCue Glucose 201 DM and HemoCue Hb 201 DM Analyzers" dated October 7, 2010, to customers notifying them of incorrect settings on error codes. The corrective action consists of an update of the configuration settings on the the affected HemoCue Glucose 201 DM and Hb 201 DM analyzers that are in the field. The affected units in the field will be corrected in accordance with a rework procedure that has been established by HemoCue AB.
Initiated
2010-10-07
Root cause
Process control
Terminated
2011-06-22
Distribution
Worldwide distribution, including USA, and Glucose 201 DM analyzers to Norway and the Netherlands; Hb 201 DM analzyers to the Netherlands, France, Ireland, Great Britain, Poland and Germany.
Recall number
Z-1848-2011
Recall status
Terminated
CDRH record ID
96751
FDA classified
2011-03-30
Recall event number
57602

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗