Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may collide with the floor. Four consignees were notified of this recall by Update Instruction XP067/12/S -- all four units have been corrected.
Source facts
Action
Siemens initiated their correction of this product in May, 2013 by sending an Safety Update Instruction XP067/12S to resolve the software error to all affected customers. All units have been corrected.
Further questions please call (610)-219-6300.
Initiated
2013-05-07
Root cause
Software design
Terminated
2013-12-24
Distribution
US Distribution including the states of CA, IN, NJ and OK.
Recall number
Z-1848-2013
Recall status
Terminated
CDRH record ID
119218
FDA classified
2013-08-01
Recall event number
65491