Z-1848-2022: Leica Surgical Operating Microscopes, ARveo 8, Part Number 10449063

Recall record Z-1848-2022 for Leica Microsystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1848-2022
Company in FDA record
Leica Microsystems, Inc.
FDA classification date
Sep 29, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

Source facts

Action
The firm sent a Field Safety Notice to its consignees on 08/26/2022 by mail. The notice described the problem and stated that A Leica Microsystems representative will be contacting the consignee's to arrange this no-cost field update of the instrument. The field update will be scheduled to occur at your convenience within the next 6 months. The notice provided the following workaround: "Advice on action to be taken by the user: " You can continue to use the system pending this field update. Based on our analysis the probability is very remote that an unacceptable patient risk will occur due to the incorrect BrightCare Plus functionality caused by the non-conforming photodiode. " However, please ensure that all users of the affected microscope are aware of the Dangers of Use as described in section 3.4 and will follow the directions in section 5.3 ( Illumination ) and section 8 ( Operation ) of the user manual when handling a surgical operating microscope see excerpts below: o 3.4 Dangers of Use : &do not set the light intensity too high. & because if &light intensity is too high, uncontrolled tissue heating may occur. o 5.3 Illumination : &begin with a low light intensity and increase it until an optimum level of illumination is achieved. o 8 Operation : &use the lowest comfortable light intensity. The notice also requested acknowledgement of receipt of the Field Safety Notice as follows: "Please acknowledge receipt of this Field Safety Notice with the digital acknowledgment form by either scanning the QR code in your mobile phone or with the following link: https://forms.office.com/r/1WPXW93ugL."
Initiated
2022-08-26
Root cause
Component change control
Distribution
US Nationwide distribution.
Recall number
Z-1848-2022
Recall status
Open, Classified
CDRH record ID
195950
FDA classified
2022-09-29
Recall event number
90818

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗