Open original FDA record ↗Recall record
- FDA identifier
- Z-1849-2011
- Company in FDA record
- Hemo Cue, Inc.
- FDA classification date
- Mar 30, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect configurations settings in software. As a result, certain error codes do not trigger appropriately. These error codes are: E01, E02 and E05 which are all related to the optronic performance of the analyzer. Possible incorrect results.
Source facts
- Action
- Hemocue AB sent Customer Notification "Notice of Medical Device Correction on HemoCue Glucose 201 DM and HemoCue Hb 201 DM Analyzers" dated October 7, 2010, to customers notifying them of incorrect settings on error codes. The corrective action consists of an update of the configuration settings on the the affected HemoCue Glucose 201 DM and Hb 201 DM analyzers that are in the field. The affected units in the field will be corrected in accordance with a rework procedure that has been established by HemoCue AB.
- Initiated
- 2010-10-07
- Root cause
- Process control
- Terminated
- 2011-06-22
- Distribution
- Worldwide distribution, including USA, and Glucose 201 DM analyzers to Norway and the Netherlands; Hb 201 DM analzyers to the Netherlands, France, Ireland, Great Britain, Poland and Germany.
- Recall number
- Z-1849-2011
- Recall status
- Terminated
- CDRH record ID
- 96752
- FDA classified
- 2011-03-30
- Recall event number
- 57602