Software Version 3.1.1 does not meet navigational accuracy.
Source facts
Action
Medtronic Navigation contacted customers by telephone on 10/12/2010 and with "Urgent Field Safety Notice" follow up letters dated 10/14 or 10/15/2010. Sites were visited by Medtronic Service Representative to complete navigational accuracy testing and either replace or correct the unit.
Initiated
2010-10-12
Root cause
Software design
Terminated
2012-06-21
Distribution
Nationwide Distribution including the states of CO, IN, and MN.
Recall number
Z-1849-2012
Recall status
Terminated
CDRH record ID
108856
FDA classified
2012-06-21
Recall event number
61741