After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the
system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).
Source facts
Action
All the consignees affected by this recall will be notified by:
The Urgent Medical Device Correction Letter/ Field Safety Notification will be
distributed to all affected users, with a description of the problem.
The Letter will also be distributed to the Varian Sales, Marketing, and Service
organizations, informing them of the issue.
Additionally, the letter will be posted to the customer support site:
http:/ /www.MyVarian.com.
Initiated
2015-05-18
Root cause
Software Design Change
Terminated
2016-09-12
Distribution
Worldwide Distribution.
Recall number
Z-1849-2015
Recall status
Terminated
CDRH record ID
137840
FDA classified
2015-06-26
Recall event number
71451