This correction is being initiated due to a software issue which results in the possibility of incorrect patient information being assigned to sample results when the following rare combination of three specific events were to occur:
1) No measurement results are generated due to an error by a urine sediment measurement.
2) The instrument is shutdown incorrectly.
3) An item rack is used for subsequent sample measurements.
Source facts
Action
The firm initiated the field correction by email on 05/03/2019. The letter stated how to recognize a product failure and the actions to be taking by the customer until the software upgrade is ready for installation.
Initiated
2019-05-03
Root cause
Software design
Terminated
2020-09-17
Distribution
US
Recall number
Z-1850-2019
Recall status
Terminated
CDRH record ID
172606
FDA record created
2019-06-20
Recall event number
82737