Open original FDA record ↗Recall record
- FDA identifier
- Z-1851-2020
- Company in FDA record
- Nobel Biocare Usa Llc
- FDA classification date
- May 4, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).
Source facts
- Action
- The firm's "Urgent Field Safety Notice", was issued on 08/28/2019 and the notices were sent to customers via email . The following actions are to be taken: 1) Identify affected devices in stock by confirming the information on the side label: 3.75x10mm, article no. 37965, lot 12129104. 2) Confirm that both the side and top label indicate 3.75x10mm. 3) Complete a Customer Acknowledgment Form and return it via email to: cs.russia@nobelbiocare.com within 5 days of receipt of the notice. 4) Return incorrectly labeled products (top label 3.75x7mm) to: Attn: Returns PFA1925 109004 Moscow Stanislavskogo str., h.21, b.3 (Correctly labeled products (top label 3.75x 10mm) do not need to be returned). 5) Returned devices will be replaced free of charge.
- Initiated
- 2019-08-28
- Root cause
- Labeling mix-ups
- Terminated
- 2021-04-28
- Distribution
- International distribution - the country Russia only.
- Recall number
- Z-1851-2020
- Recall status
- Terminated
- CDRH record ID
- 180338
- FDA classified
- 2020-05-04
- Recall event number
- 85236