Laparoscopic devices have a sterile symbol included on labels but devices are non-sterile.
Source facts
Action
On 01/20/2022, Urgent Field Safety Notices were emailed to customers who were asked to do the following:
1) Quarantine affected devices and return to the recalling firm.
2) If the devices are no longer in stock, confirm whether devices were used clinically, and if so, whether the devices were sterilized in accordance with the Instructions for Use prior to use.
3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices were transferred.
4) Complete response form.
The recalling firm provided the following email: quality@surginno.co.uk
Initiated
2022-01-20
Root cause
Process control
Distribution
US: MA
Recall number
Z-1852-2022
Recall status
Open, Classified
CDRH record ID
195860
FDA classified
2022-09-29
Recall event number
90801