Near Point Chart Arm of RT 5100 and RT-3100 Refractors may lower spontaneously with potential for patient injury.
Source facts
Action
Recall/Field Correction letters were sent to doctors on June 9, 2014 with a preventive maintenance instruction to be performed on the device by user. Also included are the necessary tools to perform actions. Customers are to complete and return the Recall Response Card. Questions are to be directed to Customer Service.
Contact Information:
Customer Service: 1-800-891-1772
E-mail: RT-5100 update@noritsu.com
Initiated
2014-06-09
Root cause
Use error
Terminated
2015-10-30
Distribution
US distribution only.
Recall number
Z-1853-2014
Recall status
Terminated
CDRH record ID
128027
FDA classified
2014-06-23
Recall event number
68485