Skip to content

Public FDA record

Z-1855-2013: List No. 14211-28; PLUM BLOOD SET; Prepierced Port, 200 Micron Filter, 110 Inch, Non-DEHP; For Blood Prod

Recall record Z-1855-2013 for Hospira Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1855-2013
Company in FDA record
Hospira Inc.
FDA classification date
Aug 5, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

It was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.

Source facts

Action
Hospira sent an Important Safety Information letter dated April 1, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to exercise EXTREME CAUTION when using Hospira blood bags and to make sure that all instructions for use are included with the blood bag and their facility's protocol for spiking blood bags are completely followed in order to minimize the possibility of puncturing the outer wall of the blood bag. Customers with issues were instructed to call Hospira at 1-800-441-4100 or email ProductComplaintsPP@hospira.com. Customers were asked to complete the attached replay formand return it to the fax number on the form. If customers distributed the product further they were instructed to notify their accounts that received the affected product and ask them to contact Stericycle at 1-866-608-3936 to receive a reply form. Customers with questions were instructed to contact Hospira at 1-800-441-4100. For questions regarding this recall call 224-212-2000.
Initiated
2013-04-01
Root cause
Device Design
Terminated
2017-05-24
Distribution
Worldwide Distribution - USA (nationwide) and Puerto Rico and Internationally to Costa Rica.
Recall number
Z-1855-2013
Recall status
Terminated
CDRH record ID
119888
FDA classified
2013-08-05
Recall event number
65689

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗