Skip to content

Public FDA record

Z-1856-2025: Glidewell HT Implant Twist Drill 1.5 x 8 mm " SKU # 70-1071-SRG0266 UDI code: +D7457071SRG02660/$$7625479

Recall record Z-1856-2025 for Prismatik Dentalcraft, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1856-2025
Company in FDA record
Prismatik Dentalcraft, Inc.
FDA classification date
May 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to packaging mix up. Package of finished devices were reported to have an incorrect finished device package inside.

Source facts

Action
On 04/14/2025, the firm sent via overnight mail an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL" Letter to customers informing them that Prismatik had shipped the device, Glidewell HT" Implant Twist Drill 1.5 x 8 mm (Device SKU #: 70-1071-SRG0266) mispackaged with an incorrect finished device, Glidewell HT" Implant Shaping Drill 5.0 x 11.5 mm (Device SKU #: 70-1071-SRG0285) packaged inside. The affected Lot Number is 6254798. Customers are instructed to: 1. Review their inventory for the product received between 03/06/2025 through 03/18/2025 with Lot Number 6254798. If they have any product remaining in inventory, collect and quarantine immediately. 2. For the products listed for this recall. discontinue use and if they have any inventory in storage, return the product to receive a credit, using the return form attached with the specific contact information to: RA.Mailbox@glidewelldental.com. 3. If completing the form: indicate the quantities that need to be returned. Sign, date, and return the form to RA.Mailbox@glidewelldental.com. Keep the quarantine product segregated; Prismatik Dentalcraft Customer Service Representative will contacting customers with further instructions. 4. Make a copy of the signed form and place it with the returning product. This will identify the product as Recall Product for proper handling and disposition. 5. If customers do not have product to be returned, place an X next to the statement, No Product to Return , sign, date and return the form to RA.Mailbox@glidewelldental.com. For questions/assistances, contact Regulatory Affairs at RA.Mailbox@glidewelldental.com or (949) 836-5479.
Initiated
2025-04-14
Root cause
Process control
Distribution
U.S Nationwide distribution in the states of KY, MA, MI, NC, VA, and WV.
Recall number
Z-1856-2025
Recall status
Open, Classified
CDRH record ID
213439
FDA classified
2025-05-28
Recall event number
96711

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗