Z-1858-2008: Physio-Control LIFEPAK 12 defibrillator/monitor. The LIFEPAK12 defibrillator/monitor series is a battery

Recall record Z-1858-2008 for Physio Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1858-2008
Company in FDA record
Physio Control, Inc.
FDA classification date
Sep 11, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The LIFEPAK12 defibrillator/monitor series (LP12) contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock.

Source facts

Action
Starting May 8, 2008, letters were hand-delivered by the firm's service representatives as they visited customer sites to perform on-site service for affected units. Customers who had not been visited by May 16th were mailed a Certified Receipt letter on May 16, 2008. The letter explained the recall and notified customers that Physio Control was conducting a Field correction to replace all affected Biphasic PCBA. The firm recommends customers continue to perform therapy cable or Standard Paddles Defibrillation Check (monthly or as needed); to have a backup device available until their device is updated; and if no backup device is available, a loaner would be provided by the firm upon request. Contact Technical Support at 1-800-442-1142, option 5 - 6:00 am to 4:00 pm (Pacific), Monday -Friday or visit www.Physio-Control-notices.com/LP12.
Initiated
2008-05-08
Root cause
Process control
Terminated
2013-09-04
Distribution
Product distributed nationwide to 5 Department of Defense accounts, 1 VA account, 5 government accounts, 49 hospitals, 223 first responders, and 1 direct account. 8 foreign accounts include: NETHERLANDS, MEXICO, JAPAN, HONG KONG, GERMANY, CANADA and AUSTRALIA.
Recall number
Z-1858-2008
Recall status
Terminated
CDRH record ID
70898
FDA classified
2008-09-11
Recall event number
48140

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗