- FDA identifier
- Z-1858-2008
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Sep 11, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The LIFEPAK12 defibrillator/monitor series (LP12) contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock.
Source facts
- Action
- Starting May 8, 2008, letters were hand-delivered by the firm's service representatives as they visited customer sites to perform on-site service for affected units. Customers who had not been visited by May 16th were mailed a Certified Receipt letter on May 16, 2008. The letter explained the recall and notified customers that Physio Control was conducting a Field correction to replace all affected Biphasic PCBA. The firm recommends customers continue to perform therapy cable or Standard Paddles Defibrillation Check (monthly or as needed); to have a backup device available until their device is updated; and if no backup device is available, a loaner would be provided by the firm upon request. Contact Technical Support at 1-800-442-1142, option 5 - 6:00 am to 4:00 pm (Pacific), Monday -Friday or visit www.Physio-Control-notices.com/LP12.
- Initiated
- 2008-05-08
- Root cause
- Process control
- Terminated
- 2013-09-04
- Distribution
- Product distributed nationwide to 5 Department of Defense accounts, 1 VA account, 5 government accounts, 49 hospitals, 223 first responders, and 1 direct account. 8 foreign accounts include: NETHERLANDS, MEXICO, JAPAN, HONG KONG, GERMANY, CANADA and AUSTRALIA.
- Recall number
- Z-1858-2008
- Recall status
- Terminated
- CDRH record ID
- 70898
- FDA classified
- 2008-09-11
- Recall event number
- 48140