Skip to content

Public FDA record

Z-1858-2022: The BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tiss

Recall record Z-1858-2022 for Hologic, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1858-2022
Company in FDA record
Hologic, Inc
FDA classification date
Sep 30, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is actually manufacturing date, and it's different from the one on the shelf box label, which is the correct expiration date.

Source facts

Action
On August 9, 2022, the recalling firm sent out recall letters with subject line "RE: Incorrect expiration date on BioZorb bioabsorbable marker removable label". Customers were instructed to return any unused product from the affected lot to Hologic. Please contact BSHCustomersupport@hologic.com with the number of unused units from Lot#22A20RL that you have. You will be provided a pre paid label to facilitate return. You will also be issued a credit on your account for the returned product. If you wish to reorder product, please place a new order via your preferred procurement method. If material has already been implanted from the affected lot and the patient chart requires updating, Hologic included corrected labels with the recall letter for that purpose.
Initiated
2022-08-09
Root cause
Incorrect or no expiration date
Distribution
US Nationwide distribution including in the states of FL, IL, MD, NJ, AZ, TX, WA, VA, OK, NC, NH, GA, CA, CT, OH, CO, MI, MO, KS, TN, ID, MS, MT, AL, OR, IA, PA, MA.
Recall number
Z-1858-2022
Recall status
Open, Classified
CDRH record ID
195599
FDA classified
2022-09-30
Recall event number
90734

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗