Open original FDA record ↗Recall record
- FDA identifier
- Z-1858-2025
- Company in FDA record
- C-RAD POSITIONING AB
- FDA classification date
- May 28, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software issue with scanning equipment that can results in the filed rotation not being applied correctly.
Source facts
- Action
- On April 22, 2025, C-RAD Positioning AB issued a Field Safety Notice to affected consignees via E-Mail. C-RAD asked consignees to take the following actions: 1. As an immediate action, the affected users are recommended to: - Do not use Site Groups for treatment plans containing non-coplanar fields. - Ungroup all existing Site Groups. 2. Ensure that all the staff operating the Catalyst family systems and the c4D software application are aware of this information. 3. All customers are requested to reply to C-RAD using the attached FSN Customer Reply Form .
- Initiated
- 2025-04-22
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, FL, WA, VA, WA, NC, IL, AR, NC and the countries of Germany, Italy, Uzbekistan, South Korea, Vietnam, Austria, Australia, China, Estonia, Belgium, Uzbekistan.
- Recall number
- Z-1858-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213672
- FDA classified
- 2025-05-28
- Recall event number
- 96791