Z-1859-2008: Model 35A Series Mobile Transporters is an ambulance stretcher for patient transport.

Recall record Z-1859-2008 for Ferno-Washington Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1859-2008
Company in FDA record
Ferno-Washington Inc
FDA classification date
Sep 11, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A component defect was found in corner castings of the firm's wheeled ambulance stretchers. Some corner castings on the stretchers were found to be over-sized. Using an over-sized cone-shaped hole corner casting with an improperly seated shank pin could put undue stress on the top screw, possibly causing the screw securing a wheel to become loose or break and the wheel to detach from the stretcher

Source facts

Action
The firm sent an 'Urgent Field Correction' notification letter to their customers, via certified mail on 5/16/2008. The letter instructs the customers to examine all inventory of Model 35A stretchers. The letter also recommends that any identified unit be removed from service until the unit has been inspected by the customer and a field correction completed, if applicable. Once the inspection of the unit(s) has been completed by the customer, they are to contact the firm immediately and inform them of the results of t their inspections. In the event that a casting fails inspection, the customer is instructed to contact Ferno's Technical Support Div. in order to arrange for a casting replacement. To report the results of their inspection(s) or to address any questions or concerns which the customer may have, contact Ferno Technical Support at 1-866-987-3776 or send an email to Quality.Products@ferno.com.
Initiated
2008-04-29
Root cause
Mixed-up of materials/components
Terminated
2012-10-30
Distribution
The recalled product was distributed to customers in the following location nationwide: AK, AL, AR, CA, CO, CT, DE, FL, GA, HI, IL, IN, KY, LA, MA, MD, ME, MI, MN, MS, NC, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WI, and WV.
Recall number
Z-1859-2008
Recall status
Terminated
CDRH record ID
70903
FDA classified
2008-09-11
Recall event number
48146

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗