Reason stated in the source
During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.
Source facts
- Action
- On 03/31/2025, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that as a result of an investigation, it has been revealed that a combination of chemical and electrical interactions that result in electrochemical degradation, compromising the structural integrity of the sensor over time. This degradation could potentially lead to water leakage into the machine. On 04/25/2025, the firm sent via FedEx and updated letter. Customers are notified and/or instructed that/to: -A field technicians will be contacting you to arrange for replacement if your unit has been impacted by this issue. -Complete and sign the attached response form to acknowledge that they have received the notification. Return the completed form to Getinge by emailing a scanned copy to recallresponses.qrc@getinge.com Public contact for questions/assistance - recalls.qrc@getinge.com
- Initiated
- 2025-03-31
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AZ, CA, CO, FL, ID, IL, IN, KS, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SD, TX, VA, WA, WI, and WY. The country of Canada.
- Recall number
- Z-1859-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213361
- FDA classified
- 2025-05-28
- Recall event number
- 96655