Z-1860-2021: Angiographic Guidewire Component: Model Number: 107042, 107044, 107447, 110003, 110004, The angiographi

Recall record Z-1860-2021 for Medtronic Vascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1860-2021
Company in FDA record
Medtronic Vascular
FDA classification date
Jun 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The component part numbers intended for shipment to a third-party, where they are further processed into kits and sterilized were instead shipped directly to customers.

Source facts

Action
In May, 2021, Medtronic issued an "URGENT: MEDICAL DEVICE RECALL" notice to customers via letter notifying them Medtronic is voluntarily recalling certain lots of Angiographic Guidewire components. In June 2021, Medtronic issued AMENDED URGENT MEDICAL DEVICE RECALL notices to customers. On The component part numbers associated with this recall are intended for shipment to a third-party, where they are further processed into kits and sterilized, but were instead shipped directly to customers. Actions: Medtronic records indicate that your facility has received one or more of the affected Angiographic Guidewire components. As a result, Medtronic requests that you immediately take the following actions: -Identify and quarantine all unused affected components identified in the customer letter. -Utilize standard hospital practice and policies to notify patients who were treated with the recalled components. -Return/Exchange all unused affected components in your inventory to Medtronic. Contact Medtronic Customer Service at 1-800-716-6700 to initiate a component return/exchange. Your local Medtronic Representative can assist you in the return of this component. Please refer to Table 2 for applicable part numbers to facilitate reordering. -Please complete the enclosed Customer Confirmation Form and email to rs.cfqfca@medtronic.com. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. If you have any questions regarding this communication, please contact your Medtronic Field Representative or call:763-526 -8583 or email: rs.cvgrecalls@medtronic.com.
Initiated
2021-05-04
Root cause
Component design/selection
Terminated
2023-01-05
Distribution
US /Domestic Distribution to states of: CA, KY, LA, MI, NE, NY, OH, OK, TX, WA, WI. OUS/Foreign to country of: Australia
Recall number
Z-1860-2021
Recall status
Terminated
CDRH record ID
187503
FDA classified
2021-06-16
Recall event number
87952

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗