Open original FDA record ↗Recall record
- FDA identifier
- Z-1861-2021
- Company in FDA record
- Skytron, Div. The KMW Group, Inc
- FDA classification date
- Jun 15, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Failure of the FS32 Monitor Brackets from the mounting block could potentially loosen over time and eventually detach from the baseplate, result in the display interface dismounting from its attachment point, causing injury to patients or staff
Source facts
- Action
- Skytron issued Urgent Medical Device Correction notice on 5/11/21. Letter states reason for recall, health risk and action to take: Identify your FS32 monitor bracket(s). Serial numbers can be found on the UDI label located on the back of the bracket. Inspect your FS32 monitor bracket(s) per the instructions on the final page of this document. Remove from service any bracket where evidence is identified that the device may fail. Skytron authorized representatives have been directed to contact you and schedule a service visit. Please contact Erin Woolf, at 616-656-2998 or ewoolf@skytron.us with any questions. ***UPDATED 12/10/21*** On November 22, 2021, the firm sent updated letters due to an expansion of the recall scope. Revision 3 models of the FS32 Monitor Bracket were added to the recall. Customer instructions remain the same.
- Initiated
- 2021-05-11
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Japan.
- Recall number
- Z-1861-2021
- Recall status
- Open, Classified
- CDRH record ID
- 187470
- FDA classified
- 2021-06-15
- Recall event number
- 87939