Open original FDA record ↗Recall record
- FDA identifier
- Z-1862-2011
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Mar 31, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Stryker Orthopaedics has become aware that the 10 mm on the bone remaining depth gage is incorrect thus allowing for the reamer to leave as little as 8.9mm (in a nominal condition) patella bone.
Source facts
- Action
- Stryker sent an Urgent Product Recall letter dated January 27, 2011, via FedEX for January 28, 2011, delivery with return receipt. The letter identified the product, the problem, and the action to be taken by the customer. Customers were insructed to fax back the attached Product Recall Acknowledgement Form within 5 days to 201-831-6069. Customers were also instructed to contact the hospitals in their territory that have the affected product to arrange return of the product. Product should be returned to the attention of: Regulatory Compliance Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey, 07430. The orange Product Remediation sticker should be attached to the return, indicating the Product Remediation # RA 2009-296. Mark the outer box with words "Product Recall." For questions regarding this recall call (201) 972-2100.
- Initiated
- 2011-01-27
- Root cause
- Device Design
- Terminated
- 2012-10-15
- Distribution
- Nationwide Distribution
- Recall number
- Z-1862-2011
- Recall status
- Terminated
- CDRH record ID
- 97584
- FDA classified
- 2011-03-31
- Recall event number
- 57852