Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field.
Source facts
Action
The firm initiated their recall by visit on 04/07/2021. VERO Clinical Educators are going (or have gone) onsite to the hospitals to physically upgrade the software to v2.2.4. The correction of upgrading the software from v2.2.3 to v2.2.4 began on Wednesday April 7, 2021. The software upgrade is expected to be completed by Friday April 23, 2021. For questions contact VERO Biotech at 1-877-337-4118 or visit the VERO Biotech website at https://www.vero-biotech.com/.
Initiated
2021-05-19
Root cause
Software design
Terminated
2023-10-12
Distribution
Distributed nationwide to TX, MD, VA, GA, FL, KS, LA, MN, and NC
Recall number
Z-1863-2021
Recall status
Terminated
CDRH record ID
186955
FDA classified
2021-06-23
Recall event number
87764