Stryker Orthopaedics became aware that when attempting to position the Triathlon Femoral Sizer, the pre-setting may slip from l (left knee) to R (right knee) and vice verse intraoperatively.
Source facts
Action
Letters were sent to all affected Stryker Branches/Agencies on May 27, 2005. Letters requested that each branch examine the inventory and hospitals locations to identify the instruments.
Contact James Young at 201-831-5110, Brian Coughlin at 210-831-5665 or Rita Intorrella at 201-831-5825 if you have questions.
Initiated
2005-05-27
Root cause
Device Design
Terminated
2008-09-24
Distribution
To Stryker branches/agencies worldwide.
Recall number
Z-1864-2008
Recall status
Terminated
CDRH record ID
70990
FDA classified
2008-09-16
Recall event number
48200