Open original FDA record ↗Recall record
- FDA identifier
- Z-1865-2011
- Company in FDA record
- Xoft Inc.
- FDA classification date
- Apr 1, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Xoft Axxent Flexishield Mini product may may shed particles identified as tungsten, and may look like suspicious calcifications on follow up scans if the product was used during Intraoperative radiation therapy.
Source facts
- Action
- Xoft, Inc. sent an Urgent Field Removal Action letter dated February 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to stop using and return all units and lots of Flexishield Mini Catalog Number: F5300 in their inventory. Customers were instructed to fill out the attached form and fax to 866-222-3404, then call Customers Service toll free directly at 877-963-8327 to make arrangements to return the product and receive a credit if applicable. Customers were asked to complete and return the form even if they did not have any units to return. For questions regarding this recall call 408-419-2300.
- Initiated
- 2011-02-03
- Root cause
- Device Design
- Terminated
- 2011-06-08
- Distribution
- Nationwide Distribution including AL, CA, CO, FL, IL, MI, MO, MS, NJ, NY, OK, PA, TN, and WI.
- Recall number
- Z-1865-2011
- Recall status
- Terminated
- CDRH record ID
- 97687
- FDA classified
- 2011-04-01
- Recall event number
- 57890