Open original FDA record ↗Recall record
- FDA identifier
- Z-1865-2012
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA classification date
- Jun 22, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mode" and "digital spot mode", and the system does not reboot.
Source facts
- Action
- On 11/20/06 two written notifications were sent to consignees addressed to the Facility Administrator and the Radiology Administrator. The notifications advised them that GE Healthcare had identified several intermittent potential safety issues that may occur with the product. Interim solutions were provided for each issue. The letter stated that GEHC OEC would contact customers and remedy the issues free of charge when a permanent solution became available. An Urgent Recall Notice, dated April 13, 2007, was also sent to customers. Customers were to contact the service team at 800-874-7378 for further information, if there were any questions or concerns.
- Initiated
- 2006-11-20
- Root cause
- Software design
- Terminated
- 2012-06-25
- Distribution
- Worldwide Distribution.
- Recall number
- Z-1865-2012
- Recall status
- Terminated
- CDRH record ID
- 109749
- FDA classified
- 2012-06-22
- Recall event number
- 62016