Skip to content

Public FDA record

Z-1865-2019: Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05

Recall record Z-1865-2019 for Thermo Fisher Scientific (Asheville) LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1865-2019
Company in FDA record
Thermo Fisher Scientific (Asheville) LLC
FDA record creation date
Jun 21, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

Source facts

Action
Thermo Fisher issued on 4/02/19 via Email notification with attachments to distributors (domestic and foreign) the Urgent Medical Recall Notification Letter with customer response sheet. Recall is to the user level.On May 23, 2019 email followed by a formal written, "Urgent Medical Device Recall Notification Letter" with customer response sheet were sent via US mail to domestic customers. Letters address health risk and action to take: customer service team is prepared to send you a replacement immediately and the necessary materials and instructions for the return of the bath you received initially to the factory. Please call representatives with your original purchase order, company name, and contact information at 866-984-3766 or email the included Recall Customer Response Form.
Initiated
2019-04-02
Root cause
Process control
Terminated
2020-07-23
Distribution
CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Recall number
Z-1865-2019
Recall status
Terminated
CDRH record ID
173093
FDA record created
2019-06-21
Recall event number
82903

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗