Open original FDA record ↗Recall record
- FDA identifier
- Z-1865-2025
- Company in FDA record
- Zyno Medical LLC
- FDA classification date
- Jun 6, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Unreleased software versions were installed on distributed devices without verification or validation.
Source facts
- Action
- Consignees received an "URGENT: MEDICAL DEVICE CORRECTION" notification dated 5/7/25 by email. The notification instructs consignees to check their stock for affected devices and to immediately stop use; recalled devices are to be returned to Zyno Medical via the provided Response Verification Form. The recall notice is to be provided to personnel at consignee locations who need to be informed and communicated to facilities to where devices were transferred to. Any questions about this recall action are to be directed via email to Frederick Lee at feedback@intuvie.com. A representative from Zyno Medical s partner company, Intuvie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action.
- Initiated
- 2025-05-07
- Root cause
- Software design (manufacturing process)
- Distribution
- US Nationwide
- Recall number
- Z-1865-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213681
- FDA classified
- 2025-06-06
- Recall event number
- 96796