Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient to contact the underlying carbon fiber structure of the Wilson frame and could lead to a pressure injury.
Source facts
Action
Letters were sent to customers and distributors on March 28, 2017 by Federal Express or UPS. Information is also posted on firm's web site. Frames having 5319-37 pads should not be used and all of the affected pads retrieved and destroyed.
Initiated
2017-03-28
Root cause
Process control
Terminated
2017-08-15
Distribution
United States, Canada, Japan, Australia, New Zealand, Great Britain, Sweden, Spain, Germany, Switzerland, Kenya, Saudi Arabia, Hong Kong, South Korea.
Recall number
Z-1867-2017
Recall status
Terminated
CDRH record ID
154562
FDA classified
2017-04-19
Recall event number
76920