Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Source facts
Action
Thermo Fisher issued on 4/02/19 via Email notification with attachments to distributors (domestic and foreign) the Urgent Medical Recall Notification Letter with customer response sheet. Recall is to the user level.On May 23, 2019 email followed by a formal written, "Urgent Medical Device Recall Notification
Letter" with customer response sheet were sent via US mail to domestic customers.
Letters address health risk and action to take: customer service team is prepared to send you a replacement immediately and the necessary materials and instructions for the return of the bath you received initially to the factory. Please call representatives with your original purchase order, company name, and contact information at 866-984-3766 or email the included Recall Customer Response Form.
Initiated
2019-04-02
Root cause
Process control
Terminated
2020-07-23
Distribution
CA, GA, KY, PA
Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Recall number
Z-1868-2019
Recall status
Terminated
CDRH record ID
173200
FDA record created
2019-06-21
Recall event number
82903