Open original FDA record ↗Recall record
- FDA identifier
- Z-1869-2013
- Company in FDA record
- SpineFrontier, Inc.
- FDA classification date
- Aug 1, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Two sections that are designated for the 8mm and 9mm screws were switched; the holes that hold the 8mm screws are designed for 9mm screws, and the holes that hold the 9mm screws are designated for 8mm screws.
Source facts
- Action
- The firm, SpineFrontier, sent an Advisory Notice entitled "URGENT May 1, 2013, PedFuse Screw Caddy Removal" dated May 1, 2013 to their consignees/customers. The notice described the product, problem and action to be taken. Consignees/customers were informed that a removal has been initiated to remove the parts from the field to ensure that all sets contain correctly marked screw caddies. Consignees/customers wre instructed to complete and return the URGENT PedFuse Screw Caddy Removal ACKNOWLEDGEMENT FORM via fax to: SpineFrontier, Inc., 978-232-3991 or email: customerservice@spinefrontier.com. Replacement caddies will be shipped upon receipt of completed acknowledgement form and RMA request for part return. Please feel free to contact SpineFrontier at 978-232-3990 with any questions.
- Initiated
- 2012-10-05
- Root cause
- Device Design
- Terminated
- 2015-07-29
- Distribution
- US distribution in states of: CO, DC, MA and TX.
- Recall number
- Z-1869-2013
- Recall status
- Terminated
- CDRH record ID
- 117585
- FDA classified
- 2013-08-01
- Recall event number
- 64987