Skip to content

Public FDA record

Z-1869-2014: MOSAIQ MOSAIQ is an oncology information system used to manage workflows for treatment planning and de

Recall record Z-1869-2014 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1869-2014
Company in FDA record
Elekta, Inc.
FDA classification date
Jun 25, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

MOSAIQ does not re-calculate the agent volume when the user changes the ordering dose (to other than what was calculated) prior to saving the order.

Source facts

Action
Elekta issued an Important Safety Notice dated May 29, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Corrective action that shall be taken: Notify the affected customers of the issue - To be completed 06/22/2014 Follow up with the affected customers for: Confirmation receipt of the LIN, AND Confirmation of installation of patch for the fix- To be completed 11/30/2014 The notice includes a confirmation which is to be returned. For further questions call (770) 670-2422.
Initiated
2014-05-29
Root cause
Software design
Terminated
2021-07-16
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, IA, KY, LA, ME, MD, MA, MI, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TX, VT, VA and WA,. and the countries of Australia, Canada, New Zealand and United Kingdom.
Recall number
Z-1869-2014
Recall status
Terminated
CDRH record ID
127501
FDA classified
2014-06-25
Recall event number
68386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗