Open original FDA record ↗Recall record
- FDA identifier
- Z-1869-2018
- Company in FDA record
- Rockwell Medical, Inc
- FDA record creation date
- May 14, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A subassembly component (dextrose bag) within this case product contained small amounts of sodium bicarbonate from a prior batch.
Source facts
- Action
- Removal notification involved one customer and two facilities. The two facilities receiving the DR-145 kit lot were contacted directly via phone, email, and issuance of a RGA to return the product. Stock retrieval was completed by 4/7/17. See attached email and RGA. Communication was already completed and stock was retrieved. Discard and/or return via RGA provided The complaint documented a stock reconciliation in the complaint and all units were accounted for. The two customer facilities involved responded directly Affected kits were reworked to remove and replace the nonconforming dextrose bag. For further questions, please call (864) 849-0530.
- Initiated
- 2018-04-07
- Root cause
- Employee error
- Terminated
- 2018-10-01
- Distribution
- US Distribution to the state of NY
- Recall number
- Z-1869-2018
- Recall status
- Terminated
- CDRH record ID
- 163799
- FDA record created
- 2018-05-14
- Recall event number
- 79878