Skip to content

Public FDA record

Z-1869-2021: HME for Tracheostomized Patients, Item Code 353S13046 - Product Usage: indicated for single use on anesth

Recall record Z-1869-2021 for Covidien, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1869-2021
Company in FDA record
Covidien, LP
FDA classification date
Jun 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The company was notified by their supplier of sterilization services for airway products in Italy of potential deviations from validated parameters for ethylene oxide sterilization.

Source facts

Action
An Urgent Medical Device Field Safety Notice dated 4/29/21 was sent to customers. Actions you should take: 1. Immediately quarantine and discontinue use of item codes with associated lot numbers listed in Attachment A. 2. If you have distributed the devices listed in Attachment A, please promptly forward the information from this letter to those recipients. Actions being taken by Medtronic: 1. Medtronic is investigating the sterilization records for affected production lots to determine whether sterilization has been achieved despite the deviations. 2. Medtronic will communicate with affected customers regarding the finding of our investigation once complete and will provide direction on what customers should do with the devices quarantined because of this Field Safety Notice. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2. An URGENT MEDICAL DEVICE WITHDRAWAL notification letter dated 7/12/21 was sent to customers. Medtronic has concluded its investigation of potential deviations in the ethylene oxide sterilization processes performed by Steril Milano, the former supplier of our sterilization services for the DAR" airway products. Medtronic analyzed the available sterilization data and conducted validation tests on production lots where data was available. For products where sterilization process data was available and testing could be conducted, we determined that sterilization of those devices was achieved. However, because there was not a consistent controlled process at the sterilization facility, we are issuing a withdrawal of the production lots of the DAR" airway products which are currently under quarantine at your facility. Required Actions: 1. Please return affected product listed on Attachment A as indicated below. All unu
Initiated
2021-04-29
Root cause
Material/Component Contamination
Distribution
Worldwide distribution - US Nationwide distribution in the states of DE, KY, NY, TX, VA, VT, and WA. Global Distribution.
Recall number
Z-1869-2021
Recall status
Open, Classified
CDRH record ID
187450
FDA classified
2021-06-16
Recall event number
87925

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗