Open original FDA record ↗Recall record
- FDA identifier
- Z-1870-2009
- Company in FDA record
- GE Medical Systems, LLC
- FDA classification date
- Sep 9, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Breathing circuit inner tube may separate from the rigid connectors under certain conditions. A separation of the inner tube from the connector may prevent adequate gas flow (O2, N2O and anesthetic agents) to the patient. A separation can also cause CO2 re-breathing. This could lead to hypoxia and/or hypercarbia and inadequate anesthesia delivery.
Source facts
- Action
- GE Healthcare sent consignees 6/5/09 an "Urgent Medical Device Correction" letter dated June 5, 2009. The letter was addressed to Health Care Administrator/Risk Manager; Chief of Anesthesia; Director of Clinical Engineering; Risk Manager. The letter described the Safety Issue, Affected Product Details, Safety Instructions (included "Recall Confirmation" form fax to 973-790-4150), Product Correction and Contact Information.
- Initiated
- 2009-06-05
- Root cause
- Process control
- Terminated
- 2014-01-16
- Distribution
- Worldwide distribution: USA, AUSTRALIA, DENMARK, FRANCE, ESTONIA, GERMANY, ITALY, JORDAN, NETHERLANDS, NORWAY, SWEDEN, and UNITED ARAB EMIRATES.
- Recall number
- Z-1870-2009
- Recall status
- Terminated
- CDRH record ID
- 83446
- FDA classified
- 2009-09-09
- Recall event number
- 52605